Supply of Ventilator: 1 General Descriptionfully Microprocessor Controlled Having Volume Cycled And Time Cycled With Volume And Pressure Preset With Invasive And Non-Invasive Modes And Facility To Monitor Respiratory Parameters Including Etco2.
2 Applicationadult As Well As Paediatric Application
3 Power Supply And Operation Modea. Electrical With Only Inbuilt Battery Backup For Minimum. 5To 6 Hrs. B. 220 V Plus Or Minus 15 Percent , 50 Hz Plus Or Minus 3 Percent. With Inbuilt Facility To Work Over A Wide Range Of Voltage Fluctuations With True Online Ups With Isolation Transformer.
4 Driving Gasa. In-Built Or External Air Source From Same Manufacturer As That Of Ventilator With Usfda Approval And Not Oem. B.It Should Either Have Facility To Connect To External Central Medical Compressed Air Line With Auto Switchover Facility Or Facility To Connect To Central Oxygen Pipeline Through High Pressure Hose And Low Pressure Oxygen Source Like O2 Cylinder Through Flow Meter, Which Is Appropriate To The Source.
B. The Compressor Based Systems Should Have Facility To Connect To External Central Medical Compressed Air Line With Auto Switchover Facility. C. Turbine Based System Must Have Both Facility To Connect To Central Oxygen Pipeline Through High Pressure Hose And Low Pressure Oxygen Source Like O2 Cylinder Through Flow Meter.As Per Tender Documents.